Fluorescence decay of Rose Bengal in water measured with the EasyLife™ V lifetime system
equipped with a 525 nm LED source.
- Lifetime range: From 100 ps to 3 µs
- Size: 15 by 13 inches (38.4 by 33.3 cm)
- Sensitivity: 400 pM fluorescein or better
- Excitation: OBB proprietary pulsed nanosecond LEDs
- Optical pulse width: 1.5 ns (typical)
- Excitation range available: 280 - 670 nm
- Emission range: 200 - 650 nm (optional to 900 nm)
- Wavelength selection: By filters
- Detection: OBB patented lifetime detector
- Typical acquisition time: 20 s (sample dependent)
- Timescale (menu selectable): Linear, arithmetic and logarithmic
- Acquisition mode: Sequential or random
- Sample holder: Single 1 x 1 cm cuvette
- Software: EasyLife™ V
- Lifetime analysis: Complete package: 1-to-4 exp, global, non-exponential, micelle kinetics, lifetime distribution (ESM, MEM), anisotropy, FRET calculator
- QuickStart DVD: Included