The Dual C-Port adapters attach directly to the C-Port of any microscope. It provides one input and two standard C-Port outputs. Both ports have access to full microscope field of view while a red diachronic mirror separates light to two output ports for above and below 620 nm.
An example of how the Dual C Port can work is when an OBB Corps. Model 4300 microscope photometer is attached to the port with under 620 nm and CCD camera to the port above 620 nm. The aperture on the photometer is down field from the red diachronic so fluorescence may be collected from a smaller area than the image viewed by the camera. The Dual C-Port adapter allows simultaneous viewing of fluorescence and bright field with trans-illuminator on and 660 nm long pass filter installed.
Other common applications include edge detection, fluorescence, electrophysiology, and any other application where view of full field during fluorescence measurements is advantageous.

Red Short Pass Dichroic Mirror,
Scan Range 450 nm to 850 nm
|

|
1 |
2 (1 front and 1 side) |
4 |
1 |
C-port (input/output) |
Red Short-Pass Beam splitter, Cutoff
Wavelength = 620 nm |
 |